Common faults and causes of electric distilled water

Common faults and causes of electric distilled water

The scope of use of electric distilled water :

The electrothermal distilled water device adopts tap water heating mode to obtain distilled water by circulating water condensation for use by medical and health research and laboratory units.

Common faults of electric distilled water are as follows:

A blown fuse <br> <br> this failure occurs more, certainly at the short.

After the power is cut off, the electric heating wire of each electric heating pipe can be used to measure whether the electric heating wire is grounded. If it is grounded, the two ends of the electric heating wire can be removed. This can be used as an emergency measure. The distiller is still available, but the power is reduced, and the water output is corresponding. Reduce or replace it. For example, the 20 liter/hour electric distilled water is 6 sets of 3 sets, which are used in parallel.

Sometimes the electric heating wire is not grounded when the power is off, and the grounding is burned when the power is turned on for a while.

At this time, you can check whether the scale is thick, if it is thick, it can be removed; if it is not thick, it may be caused by the thermal expansion of the electric heating wire of a certain electric heating tube.

Second, there is noise on the power switch and spark <br> <br> a noise on the power switch and the spark or short circuit the electrical circuit described not in contact, resulting in too much current.

The cause of excessive current is not only the grounding of the heating wire in the electric heating tube, but also the leakage between the electric heating tube and the evaporation pot, causing a short circuit at the power connector, or the ignition of the power connector due to moisture, resulting in the resistance of the wiring insulation board. Become smaller.

Third, reduce the amount of distilled water used <br> <br> with increasing time, also the scale of electric heating is gradually increased, the heat of the heating tube is not easy to spread, gradually reducing the amount of water distilled out, when the available physical or chemical methods Clear Scale.

If the amount of water is significantly reduced, there must be a heating wire that is blown, causing a decrease in power and a decrease in the amount of water. A multimeter can be used to detect the burnt electric heating tube and replace it.

After replacement, the water should be drained to observe whether it is leaking or seeping water. It can be energized only after there is no water seepage.

Disposable Gloves

Disposable gloves have also played a key role in the epidemic, and can be said to be indispensable anti-epidemic supplies.

Our disposable gloves are divided into Pure Nitrile Gloves,Mixed Nitrile Gloves and PVC Glove.

Disposable inspection Pure nitrile gloves,Protect your hands from infection.

Application scenarios:medical examination,food production processing,personal protection,daily household cleaning.

Function: Anti-scratch, Heat Insulation, Anti-Static

Disposable Gloves

Disposable Pvc Gloves ,Protection Safety Gloves ,Medical Gloves,Disposable Medical Gloves

Dongguan Maimeng Culture and Creative Co. Ltd , http://www.factorydirectmask.com